MET ONE 3400 portable air particle counter
The MET ONE 3400 portable air particle counter is designed specifically for ISO14644 cleanroom applications. Organize your routine cleanroom monitoring with embedded location and sample recipe management. The MET ONE 3400 portable air particle counter classification wizard assures compliance during periodic cleanroom certification.
Features:
Hot-swappable batteries –100% instrument availability
Hot-swappable batteries eliminate the need to stop measuring while your instrument charges or for powering down, zero counting and powering back up
Test Wizard for Easier Compliance - organize your cleanroom
Save time and eliminate errors when monitoring your cleanroom environments
Flexible communications - easy setup and data transfer
Reduce your paperwork with wireless, Ethernet and serial communication options to MET ONE Data Management or your own software solutions
Intuitive User Interface
Easily set up sampling plans by area, location and operating parameter, saving operator time. Eliminate setup errors and save time by exporting setup configuration to USB memory stick and deploy your sampling plan across your entire particle counting fleet!
Multiple User Login - assures compliance
Password protected multi-level user access to instrument provides data security for 21 CFR Part 11 compliance
ISO 21501-4 reliability and accuracy
Designed for exceptional particle counting performance providing ISO 21501-4 compliant unit-to-unit accuracy and reproducibility
Sequential sampling - manifold system
Low cost site monitoring solutions can be deployed with the MET ONE Model 3432 manifold. Sample sequentially from up to 32 locations with a single MET ONE Model 3400 airborne particle counter.
Specification
The MET ONE 3400 portable air particle counter is designed specifically for ISO14644 cleanroom applications. Organize your routine cleanroom monitoring with embedded location and sample recipe management. The MET ONE 3400 portable air particle counter classification wizard assures compliance during periodic cleanroom certification.
Features:
Hot-swappable batteries –100% instrument availability
Hot-swappable batteries eliminate the need to stop measuring while your instrument charges or for powering down, zero counting and powering back up
Test Wizard for Easier Compliance - organize your cleanroom
Save time and eliminate errors when monitoring your cleanroom environments
Flexible communications - easy setup and data transfer
Reduce your paperwork with wireless, Ethernet and serial communication options to MET ONE Data Management or your own software solutions
Intuitive User Interface
Easily set up sampling plans by area, location and operating parameter, saving operator time. Eliminate setup errors and save time by exporting setup configuration to USB memory stick and deploy your sampling plan across your entire particle counting fleet!
Multiple User Login - assures compliance
Password protected multi-level user access to instrument provides data security for 21 CFR Part 11 compliance
ISO 21501-4 reliability and accuracy
Designed for exceptional particle counting performance providing ISO 21501-4 compliant unit-to-unit accuracy and reproducibility
Sequential sampling - manifold system
Low cost site monitoring solutions can be deployed with the MET ONE Model 3432 manifold. Sample sequentially from up to 32 locations with a single MET ONE Model 3400 airborne particle counter.
Specification
Standard Size Channels | 3413 and 3423 0.3, 0.5, 1.0, 3.0, 5.0, 10.0 µm |
3415 and 3425 0.5, 1.0, 2.0, 3.0, 5.0, 10.0 or 0.5, 1.0, 2.0, 3.0, 5.0, 25.0* μm | |
3445 0.5, 1.0, 2.0, 3.0, 5.0, 10.0um | |
Flow Rate | 3413 and 3415 28.3 LPM (1.0 CFM) ± 5% (default factory setting) |
3423 and 3425 50 LPM (1.77 CFM) ± 5% (default factory setting) | |
3445 100LPM (3.53 CFM) ± 5% (default factory setting) | |
Zero Count | Conforms to JIS B9921. 1 count or less in 5 minutes, 95% confidence level |
Coincidence Loss | 28.3 LPM (1.0 CFM) 5% at 14,126,000 particles/m3 (400,000 particles/ft3) 50 LPM (1.77 CFM) 5% at 4,000,000 particles/m3 (113,266 particles/ft3) 100 LPM (3.53 CFM) 10% at 10,000,000 particles/m3 (283,166 particles/ft3) |
Counting Efficiency | 3413 and 3423 50% ± 20% for 0.3um (100% ± 10% at 1.5 times the minimum sensitivity), fully complies with ISO 21501-4 |
3415, 3425 and 3445 50% ± 20% for 0.5um (100% ± 10% at 1.5 times the minimum sensitivity), fully complies with ISO 21501-4 | |
Light Source | Long Life Laser™ diode with 10-year Mean Time To Failure (MTTF) |
Pump Type | Air vacuum, rated for continuous use |
Count Display | Color 1/4 VGA TFT Touch Screen |
Interface | Windows CE®-based |
Sample/Hold/Delay Times | 1 second to 23 hrs 59 minutes 59 seconds |
Count Alarms | 1 to 9,999,999 counts |
Count Cycles | Up to 100 while in automatic mode |
Location Labels | 0 to 999, appears on printout |
Data Storage | 5,000 samples, scrollable on Historical Data review screen |
Communication Interfaces | RS232, RS485, 802.11 b/g Wireless, 802.3 Ethernet, USB Client (Version 1.1), USB Host (Version 1.1) |
Communication Protocol | Modbus TCP, Modbus RTU, Serial FX |
Battery Type | Lithium ion smart battery; rechargeable, ejectable and hot-swappable |
Operating Time (Battery) | 28.3 LPM (1.0 CFM) 6 hours** 50 LPM (1.77 CFM) 6 hours** 100 LPM (3.53 CFM) Up to 3 hours** |
Battery Recharge Time | 6.75 hours minimum, 10 hours maximum |
Power | 24 VDC 3.2A with 100~240 VAC 50/60 Hz adapter (included in ship kit) |
Size, Dimensions | 31.8 W x 25.4 D x 20.3 H cm (12.5 x 10 x 8 inches) |
Weight (without battery) | 3413 and 3415 7.55 kg (16.6 lbs) 3423 and 3425 8.33 kg (18.3 lbs) 3445 8.65 kg (19.0 lbs) |
Enclosure Material | Stainless Steel |
Environment | Operating 0 to 40° C (32 to 104° F), 10 to 90% relative humidity, non-condensing Storage -40 to 50° C (-40 to 122° F), 0 to 98% relative humidity, non-condensing |
* Channel selections can be selected at time of order in a range from 0.3 μ to 25 μ. However 0.3 μ and 25 μ cannot be configured together. | |
** Battery life is estimated with these conditions: The 1.77 cfm unit sampling for 20 minutes (1 m3) sample, print record, a 5-minute hold time (simulating move to new location), then repeat this cycle. Backlight sleep time is set at 2 minutes. The 1.0 cfm unit, sample time is set at one minute, print record, a 1-minute hold time (simulating continuous sampling mode), then repeat this cycle. Backlight would remain on at all times. |