MET ONE 2400 portable air particle counter
The MET ONE 2400 portable air particle counter has years of field-proven performance continues to make this particle counter the instrument of choice for many contamination control professionals. Long Life Laser technology extends the average service life of the instrument’s laser to over ten years.
Features:
Industry leading quality and reliability
Long Life Laser™ diode reduces the need for replacement parts and the overall cost of ownership.
Sequential Sampling - Manifold System
Low cost site monitoring solutions can be deployed with the Met One Model A3432 manifold. Sample sequentially from up to 32 locations with a single MET ONE Model 2400 Series air particle counter.
Easy access to information
The Met One 2400 is extremely easy to operate and the bright LED display allows easy reading of data. The built-in printer allows for immediate or delayed printing of results of up to 400 stored samples.
Specification
Applications
· Cleanroom environmental monitoring
· Cleanroom classification
· Cleanroom validation
· Cleanroom particulate monitoring
· Aseptic pharmaceutical production
· Continuous monitoring of manufacturing processes
· Process troubleshooting particle contamination sources
· HEPA filter leak detection
· High pressure gas monitoring with simple gas selection menu and automatic sample flow ad
The MET ONE 2400 portable air particle counter has years of field-proven performance continues to make this particle counter the instrument of choice for many contamination control professionals. Long Life Laser technology extends the average service life of the instrument’s laser to over ten years.
Features:
Industry leading quality and reliability
Long Life Laser™ diode reduces the need for replacement parts and the overall cost of ownership.
Sequential Sampling - Manifold System
Low cost site monitoring solutions can be deployed with the Met One Model A3432 manifold. Sample sequentially from up to 32 locations with a single MET ONE Model 2400 Series air particle counter.
Easy access to information
The Met One 2400 is extremely easy to operate and the bright LED display allows easy reading of data. The built-in printer allows for immediate or delayed printing of results of up to 400 stored samples.
Specification
Size Channels (μm) | 2 ch. 0.3, 0.5 |
4 ch. 0.3, 0.5, 1.0, 5.0 | |
5 ch. 0.3. 0.5, 1.0, 5.0, 10.0 | |
6 ch. 0.3, 0.5, 1.0, 3.0, 5.0. 10.0 | |
Flow Rate | 1.0 CFM (28.3 LPM) ±10% |
Flow Control | Electronic meter with rear panel adjustment |
Zero Count | Less than 1 count every 5 minutes |
Display | Red LED, 7 digits |
Light Source | Laser diode (10-year MTTF) |
Outputs | Built-in printer; RS-232/RS-485 serial interface |
Sample and Hold Times | 1 second to 24 hours |
Alarm Counts | 0 to 9,999,999 |
Location Labels | 0 to 999, appear on printout |
Count Data | Total counts, counts/ft3, counts/m3 |
Data Storage | 400 samples |
Coincidence Loss | Less than 5% at 400,000 counts/ft3 |
FS209E/ISO 14644-1 Calculations | Mean of averages; standard deviation; standard error; UCL |
Power | 100 VAC, 115 VAC or 230 VAC; 110 watts |
Dimensions | 28.4 W x 15.2 H x 45.7 D cm (11.2” x 6.0” x 18”) |
Weight | 10.9 kg (24 lbs) |
Environment | Operating: 12 to 29°C (55 to 84°F), 20 to 95% relative humidity, non-condensing |
Storage: -23 to 70°C (-10 to 160°F), up to 98% relative humidity, non-condensing |
Applications
· Cleanroom environmental monitoring
· Cleanroom classification
· Cleanroom validation
· Cleanroom particulate monitoring
· Aseptic pharmaceutical production
· Continuous monitoring of manufacturing processes
· Process troubleshooting particle contamination sources
· HEPA filter leak detection
· High pressure gas monitoring with simple gas selection menu and automatic sample flow ad