MET ONE 2400 Particle Counter USA

MET ONE 2400 portable air particle counter

The MET ONE 2400 portable air particle counter has years of field-proven performance continues to make this particle counter the instrument of choice for many contamination control professionals.  Long Life Laser technology extends the average service life of the instrument’s laser to over ten years.

Features:

Industry leading quality and reliability

Long Life Laser™ diode reduces the need for replacement parts and the overall cost of ownership.

Sequential Sampling - Manifold System

Low cost site monitoring solutions can be deployed with the Met One Model A3432 manifold. Sample sequentially from up to 32 locations with a single MET ONE Model 2400 Series air particle counter.

Easy access to information

The Met One 2400 is extremely easy to operate and the bright LED display allows easy reading of data. The built-in printer allows for immediate or delayed printing of results of up to 400 stored samples.

Specification


Size Channels (μm) 2 ch.  0.3, 0.5
4 ch.  0.3, 0.5, 1.0, 5.0
5 ch.  0.3. 0.5, 1.0, 5.0, 10.0
6 ch.  0.3, 0.5, 1.0, 3.0, 5.0. 10.0
 
Flow Rate 1.0 CFM (28.3 LPM) ±10%
Flow Control Electronic meter with rear panel adjustment
Zero Count Less than 1 count every 5 minutes
Display Red LED, 7 digits
Light Source Laser diode (10-year MTTF)
Outputs Built-in printer; RS-232/RS-485 serial interface
Sample and Hold Times 1 second to 24 hours
Alarm Counts 0 to 9,999,999
Location Labels 0 to 999, appear on printout
Count Data Total counts, counts/ft3, counts/m3
Data Storage 400 samples
Coincidence Loss Less than 5% at 400,000 counts/ft3
FS209E/ISO 14644-1 Calculations Mean of averages; standard deviation; standard error; UCL
Power 100 VAC, 115 VAC or 230 VAC; 110 watts
Dimensions 28.4 W x 15.2 H x 45.7 D cm (11.2” x 6.0” x 18”)
Weight 10.9 kg (24 lbs)
Environment Operating:  12 to 29°C (55 to 84°F), 20 to 95% relative humidity, non-condensing
 
Storage: -23 to 70°C (-10 to 160°F), up to 98% relative humidity, non-condensing

Applications 

·    Cleanroom environmental monitoring
·    Cleanroom classification
·    Cleanroom validation
·    Cleanroom particulate monitoring
·    Aseptic pharmaceutical production
·    Continuous monitoring of manufacturing processes
·    Process troubleshooting particle contamination sources
·    HEPA filter leak detection
·    High pressure gas monitoring with simple gas selection menu and automatic sample flow ad